FDA 483 - iRhythm Technologies, Inc. - August 12, 2022
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
An FDA inspection of iRhythm Technologies, Inc. identified significant deficiencies in their quality system related to the Zio AT Cardiac Monitoring System. The firm failed to adequately establish corrective and preventive actions, resulting in delayed clinician notifications for life-threatening arrhythmias and a patient death. Additional issues included a failure to submit a Medical Device Report for the death, inadequate validation of manufacturing processes for re-used components, and improper handling of customer complaints.
ID · 1f6b9f5b-915b-46ab-9ce1-83f72a64f8ce