FDA 483 - Mannkind Corporation - October 03, 2024
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
An FDA inspection of Mannkind Corporation, a manufacturer of the V-Go wearable insulin delivery device, identified significant deficiencies in their quality system. Observations included inadequate statistical rationale for sampling, insufficient design change verification for water resistance, and a failure to validate the device's conformity to user needs regarding adhesion. Additionally, the firm's corrective and preventive action procedures were not fully implemented or verified, and complaint investigations were found to be inadequate, leading to potential patient safety risks.