483
Medical Solutions International, Inc.FDA 483 - Medical Solutions International, Inc. - April 15, 2024
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
Record Details
Medical Solutions International, Inc. in Merriam, KS, a manufacturer of Class II medical devices, was cited for significant deficiencies in its design control procedures. The inspection revealed inadequate establishment of procedures for design validation, incomplete risk analysis, and insufficient procedures for both design verification and design change. All four observations were noted as repeat findings from a previous FDA inspection.
Open in Dashboard
ID · 9556220c-0b6f-4bf7-a372-1ffd4062e9a5