FDA 483 - Meridian Medical Technologies, LLC a Subsidiary of Kindeva Drug Delivery, L.P. - December 21, 2019
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
An FDA inspection of Meridian Medical Technologies, Inc. identified significant deficiencies across its quality system, primarily impacting the EpiPen autoinjector. The firm failed to adequately establish procedures for corrective and preventive actions, design verification, design validation, and complaint handling. These issues led to concerns about device failures like spontaneous activation and an inability to activate, indicating a lack of robust controls over product design, manufacturing, and post-market surveillance.
- Inspection Date
- December 21, 2019
- Product Type
- Device
ID · d80db876-4516-4cc3-96c1-a0958fbe723b