483
MIRA, Inc.FDA 483 - MIRA, Inc. - June 12, 2024
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
Record Details
MIRA, Inc. in Uxbridge, MA, was cited with three observations during an FDA inspection, indicating deficiencies in its quality system. The firm failed to adequately establish procedures for equipment calibration, resulting in the use of an out-of-calibration test gauge for medical device manufacturing. Additionally, MIRA, Inc. did not conduct internal quality audits or management reviews in 2023, contrary to its own established procedures.