Richardson Imaging System in LaFox, IL was inspected from May 23-26, 2022, regarding its manufacturing of EPRC X-Ray tube housing assemblies. The inspection revealed a significant deficiency in the firm's complaint handling procedures. Specifically, complaint records lacked crucial details, hindering the adequate evaluation of potential medical device reportable events.