FDA 483 - Sterilmed, Inc. - May 22, 2024
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
An FDA inspection of Sterilmed, Inc., a reprocessor of single-use medical devices in Plymouth, MN, revealed significant deficiencies in its quality system. The firm failed to adequately control production processes, leading to reprocessed devices not conforming to specifications, including issues with biological foreign material and cracked distal tips. Additionally, the inspection found that corrective and preventive actions were not properly established, and risk analysis documentation was inadequate in identifying failure modes and causes.
- Person
ID · 5fe2c4d8-d89d-499d-9613-153348e8358d