FDA 483 - SUREPULSE MEDICAL LTD - August 01, 2024
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
SUREPULSE MEDICAL LTD in Nottingham, United Kingdom, was inspected regarding its VS Newborn Heart Rate Monitor system. The inspection revealed significant deficiencies across its quality system, including failures to report field corrections to the FDA and inadequate corrective and preventive actions. Further issues were noted with acceptance activities, process validation, supplier control, risk analysis, and complaint handling, including evaluation for Medical Device Reporting.
ID · 8a1892bc-17a3-41e8-a890-378ed8e3b583