FDA 483 - Vesta Inc. - October 20, 2017
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
During an inspection, medical device manufacturer VESTA in Franklin, WI, was cited for significant deficiencies in process validation for its sterilizers, including failure to follow established protocols and a lack of objective evidence for critical steps. Additionally, the firm lacked adequate procedures for preventing contamination, specifically concerning environmental monitoring frequency and clean room sampling. These issues indicate a failure to ensure product quality and adherence to quality system requirements.
ID · 6bc1cba3-4bf6-4a39-ad03-3fe5bb3dc991