483
WALDEMAR LINK GMBH & CO. KGFDA 483 - WALDEMAR LINK GMBH & CO. KG - August 22, 2019
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
Record Details
An FDA inspection of WALDEMAR LINK GMBH & CO. KG, a medical device manufacturer in Norderstedt, Germany, revealed two significant observations. The firm was cited for inadequate process validation, specifically concerning the pouch sealing of an ankle prosthetic device, where observed sealing temperatures deviated from validated parameters. Additionally, the inspection found deficiencies in the documentation and timely closure of corrective and preventive actions.