FDA 483 - ZIIP INC - January 12, 2026
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
ZIIP INC, a manufacturer of the ZIIP HALO device in Pleasant Hill, CA, was cited for significant deficiencies in its quality system during an FDA inspection. Observations included inadequate complaint handling, ineffective corrective and preventive actions, and failures in design control processes, including verification, validation, and change management. The firm also failed to adequately assess risks, submit required MDRs for serious malfunctions, and maintain a device master record.
- Person
ID · 271b8083-f3b3-47ac-9fb0-0ea892dfa12d