FDA WARNING_LETTER - 77 Elektronika Kft. - March 12, 2015
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
On October 8, 2015, the FDA issued a Warning Letter to 77 Elektronika Kft. following an inspection from March 9-12, 2015, at their Budapest, Hungary facility. The inspection revealed that the firm's Urisys 1100 Urine Analyzer, a Class II medical device, was adulterated under 21 U.S.C. § 351(h) due to non-conformity with Quality System (QS) regulation (21 CFR Part 820).
Key violations included: 1. **Failure to validate processes (21 CFR 820.75(a)):** The firm did not validate critical manufacturing processes for the Urisys 1100, including (b)(4) and (b)(4) processes. Their response, citing 100% functional testing, was deemed inadequate as it lacked evidence of systemic review, defined acceptance criteria, statistical rationale for sampling, and employee training plans. 2. **Failure to monitor and control process parameters (21 CFR 820.75(b)):** Procedures for monitoring and controlling parameters for validated processes like (b)(4) were not established. The firm's response was inadequate for not providing evidence of a systemic retrospective review, defining monitoring procedures, or training employees. 3. **Failure to control purchased product and