FDA WARNING_LETTER - Aju Pharm Co., Ltd. - January 23, 2025
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
On May 21, 2025, the FDA issued a Warning Letter to Aju Pharm Co., Ltd. following an inspection from January 20-23, 2025, in Seongnam-Si, Gyeonggi, South Korea. The inspection found that the firm's Class II Fixone Hybrid Knotless Anchor and Fixone Biocomposite Anchor devices were adulterated under 21 U.S.C. § 351(h) due to non-conformity with Quality System regulation (21 CFR Part 820) and failure to establish and maintain medical device report procedures (21 CFR § 803.17).
Specific violations included: 1. **Failure to adequately establish procedures for validating device design (21 CFR 820.30(g))**: The firm distributed devices since 2023 without performing design validation activities, and their corrective action plan was deemed inadequate for lacking a systemic approach and patient risk assessment. 2. **Failure to adequately establish and maintain CAPA procedures (21 CFR 820.100(a))**: The firm did not initiate CAPAs for identified suture anchor issues from 2022-2023 complaint data. Their response was inadequate for not including a systemic approach, retrospective review, or investigation into the cause of nonconformities. 3. **Failure to