FDA WARNING_LETTER - Atom Medical Corporation - September 20, 2013
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
On March 28, 2014, the FDA issued a Warning Letter to Atom Medical Corporation following an inspection from September 17-20, 2013, in Saitama City, Japan. The inspection revealed that the firm's infant incubators, warmers, transport incubators, infusion pumps, infant resuscitation units, and gynecological/obstetrics equipment are adulterated under 21 U.S.C. § 351(h) due to non-conformity with Quality System (QS) regulation (21 CFR Part 820).
Key violations include: 1. **Failure to establish and maintain CAPA procedures (21 CFR 820.100(a)(3)):** The firm did not take preventive actions for recurring serious issues like electrical current leakage in Dual Incu I incubators, and failed to perform retrospective reviews or provide adequate employee training. 2. **Failure to establish and maintain complaint handling procedures (21 CFR 820.198(a)):** Complaint procedures (AS-1402) and forms did not ensure evaluation for Medical Device Reporting (MDR) reportability (21 CFR Part 803). Retrospective review and employee training were also inadequate. 3. **Failure to evaluate and select suppliers (21 CFR 820.50(a)(1)):** The firm
ID · 39ba6293-d8e4-47aa-a2c9-832d6603ae7c
Full citation text and observation details available on the Dashboard.