WARNING LETTER
•AZ Instrument (Shen Zhen) Co., Ltd.•November 11, 2010

FDA WARNING_LETTER - AZ Instrument (Shen Zhen) Co., Ltd. - November 11, 2010

View on Dashboard

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

ChatGPT

Claude

Perplexity

Grok

Copilot

Record Details

An FDA inspection of AZ Instrument (Shen Zhen) Co., Ltd. in Guangdong, China, from November 8-11, 2010, revealed that the firm's Glucose Test Strips, classified as medical devices, are adulterated. The manufacturing methods, facilities, and controls do not conform to the Current Good Manufacturing Practice (CGMP) requirements of the Quality System (QS) regulation (21 CFR Part 820). The firm's response to the FDA 483 was not reviewed due to late submission. Key violations include: 1) Failure to adequately validate processes where results cannot be fully verified by subsequent inspection and test, as required by 21 CFR 820.75. Specifically, the critical (b)(4) process for glucose test strips lacked full validation to establish capability and reproducibility. 2) Failure to establish and maintain adequate procedures to control nonconforming product, as required by 21 CFR 820.90(a). The firm's procedures did not detail how in-process nonconformities would be documented for trending or define thresholds for investigation. The FDA requires prompt corrective action. Failure to address these violations may lead to regulatory actions, including detention of devices entering the U.S., impact on federal contracts, and denial of premarket approvals for Class III devices and Certificates to Foreign Governments. The firm must notify the FDA within 15 working days with a detailed plan, including documentation and a timetable for corrections, and investigate the root causes of these systemic quality assurance issues.

Company
AZ Instrument (Shen Zhen) Co., Ltd.
Inspection Date
November 11, 2010
Product Type
Devices
Office
Center for Devices and Radiological Health
Person
  • Alberto Gutierrez (Director)
Open in Dashboard

ID · 003b3c69-8875-450d-bae5-364ebbe9e4b9

Violation Codes6
21 U.S.C. 321(h)21 U.S.C. 351(h)21 CFR 82021 CFR 820.7521 CFR 820.90(a)21 U.S.C. 381(a)

Full citation text and observation details available on the Dashboard.

footer
Global Key Solutions
Regulatory Data Transformed
Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@globalkeysolutions.net
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026, Global Key Solutions All rights reserved.