WARNING LETTER
•Cardionics SA•July 29, 2015

FDA WARNING_LETTER - Cardionics SA - July 29, 2015

View on Dashboard

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

ChatGPT

Claude

Perplexity

Grok

Copilot

Record Details

On October 16, 2015, the FDA issued a Warning Letter to Cardionics SA following a July 27-29, 2015 inspection in Brussels, Belgium. The inspection revealed that their CarTouch Portable ECG Electrocardiograph Device is adulterated under 21 U.S.C. § 351(h) due to non-conformity with Quality System (QS) regulation (21 CFR Part 820) and misbranded under 21 U.S.C. § 352(t)(2) for failing to furnish required information under 21 CFR Part 803 (Medical Device Reporting).

Key violations include: 1. **CAPA (21 CFR 820.100(a))**: Failure to adequately establish and maintain CAPA procedures, specifically lacking analysis of quality data sources, use of statistical methodology for recurring problems, and verification/validation of CAPA effectiveness. Eleven CAPAs initiated since 2013 were not verified/validated. The firm's response was inadequate, lacking a retrospective review of design files. 2. **Complaint Handling (21 CFR 820.198(a))**: Failure to adequately establish and maintain procedures for receiving, reviewing, and evaluating complaints. The procedure did not ensure evaluation for MDR reportability, clear identification of MDR events, or complete complaint investigation

Company
Cardionics SA
Inspection Date
July 29, 2015
Product Type
Devices
Office
Center for Devices and Radiological Health
People
  • Sean M. Boyd (Office Director)
  • Jeff Morgan
Open in Dashboard

ID · 76f5e996-c85d-40ae-b948-1d111a248f9f

Violation Codes10
21 U.S.C. 321(h)21 U.S.C. 351(h)21 CFR 82021 CFR 820.100(a)21 CFR 820.198(a)21 CFR 80321 CFR 820.30(g)21 CFR 820.18121 CFR 820.18421 U.S.C. 352(t)(2)

Full citation text and observation details available on the Dashboard.

footer
Global Key Solutions
Regulatory Data Transformed
Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@globalkeysolutions.net
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026, Global Key Solutions All rights reserved.