•Cepheid Ab•March 31, 2015

FDA WARNING_LETTER - Cepheid Ab - March 31, 2015

View on Dashboard

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

ChatGPT

Claude

Perplexity

Grok

Copilot

Record Details

On July 23, 2015, the FDA issued a Warning Letter to Cepheid Ab in Solna, Sweden, following an inspection from March 30-31, 2015. The inspection revealed that the firm's Xpert Norovirus devices were adulterated under section 501(h) of the Federal Food, Drug, and Cosmetic Act, as manufacturing methods, facilities, or controls did not conform to the Quality System regulation (21 CFR Part 820).

Key violations included: 1. **Failure to establish and maintain design change procedures (21 CFR 820.30(i))**: The firm lacked procedures for identifying, documenting, validating, reviewing, and approving pre-production design changes for Xpert Norovirus. Post-production design change verification procedures were also undefined. The firm's response was inadequate, lacking documentation of corrective action implementation, a retrospective review of design changes, employee training, and systemic corrective action for design control procedures. 2. **Failure to establish and maintain CAPA procedures (21 CFR 820.100(a))**: The CAPA procedure did not require verification/validation to ensure no adverse effect on finished devices or dissemination of quality problem information to responsible personnel. The response was inadequate, lacking documentation of retrospective CAPA reviews for effectiveness, adverse effects, information dissemination, and completion, as well as evidence of training.

Company
Cepheid Ab
Inspection Date
March 31, 2015
Product Type
Devices
Office
Center for Devices and Radiological Health
Person
  • Alberto Gutierrez (Director)
Open in Dashboard

ID · 8b64eb1b-d0ce-4ab2-80d7-e2a31f233a06

Violation Codes7
21 U.S.C. 321(h)21 U.S.C. 351(h)21 CFR 82021 CFR 820.30(i)21 CFR 820.100(a)21 CFR 820.2221 CFR 820.50

Full citation text and observation details available on the Dashboard.

footer
Global Key Solutions
Regulatory Data Transformed
Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@globalkeysolutions.net
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026, Global Key Solutions All rights reserved.