FDA WARNING_LETTER - Dongguan Jianwei Electronics Products Co., Ltd. - June 06, 2014
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
On October 16, 2014, the FDA issued a Warning Letter to Dongguan Jianwei Electronics Products Co., Ltd. following a June 3-6, 2014 inspection. The inspection revealed that the firm's Class II digital infrared (IR) ear thermometers and digital oral thermometers were adulterated under 21 U.S.C. § 351(h) due to non-conformity with Quality System (QS) regulation (21 CFR Part 820).
Key violations included: 1. **Failure to establish and maintain CAPA procedures (21 CFR 820.100(a)):** Procedures lacked requirements for verifying/validating CAPAs, implementing/recording changes, and disseminating information. Specific complaint investigations lacked documented dates/results, and a corrective action was not implemented before closure. The firm's response was inadequate as it did not address reviewing previous CAPA records. 2. **Failure to establish and maintain complaint handling procedures (21 CFR 820.198(a)):** Procedures did not ensure uniform/timely processing (lacking timeframes or complaint number recording) or MDR reportability evaluation. A specific complaint lacked documentation for the decision not to investigate. The firm's response was inadequate, failing to indicate a retrospective review of complaints. 3. **Failure to establish and maintain rework procedures (21 CFR
- Inspection Date
- June 6, 2014
- Product Type
- Devices
ID · fcc837fa-86a0-4d18-b57a-9b821c5497e7
Full citation text and observation details available on the Dashboard.