The FDA issued a Warning Letter to a manufacturer of silicone facial and body contouring implants following an inspection from July 19 to August 5, 2021. The inspection revealed the firm's devices are adulterated under 21 U.S.C. § 351(h) due to non-conformity with Quality System (QS) regulation (21 CFR Part 820) and misbranded under 21 U.S.C. § 352(t)(2) for failing to furnish required information under 21 U.S.C. § 360i and 21 CFR Part 803 (Medical Device Reporting).

Key violations include: 1. **Process Validation (21 CFR 820.75(a)):** Failure to validate sterilization, silicone curing, and packaging sealing processes. The firm's responses were inadequate, lacking validation plans and interim actions. 2. **Complaint Files (21 CFR 820.198(a)):** Inadequate documentation of complaint details, investigation results, and justifications for not investigating. Redocumentation of complaints was deemed insufficient. 3. **Corrective and Preventive Action (CAPA) (21 CFR 820.100(a)):** Failure to adequately implement CAPA procedures, specifically regarding assessing the impact of process nonconformances.

Company
dsaart, llc.
Inspection Date
August 5, 2021
Product Type
Devices
Office
Center for Devices and Radiological Health
People
  • Binita S. Ashar (Director)
  • Shari J. Shambaugh (Program Division Director)
Open in Dashboard

ID · dcfd4f26-39f3-4620-ae23-a25ec5f61a94

Violation Codes10
21 U.S.C. 321(h)21 U.S.C. 351(h)21 CFR 82021 CFR 820.75(a)21 CFR 820.198(a)21 CFR 820.198(e)21 CFR 820.198(b)21 CFR 820.100(a)21 CFR 820.90(a)21 CFR 820.30(g)

Full citation text and observation details available on the Dashboard.

footer
Global Key Solutions
Regulatory Data Transformed
Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@globalkeysolutions.net
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026, Global Key Solutions All rights reserved.