WARNING LETTER
•EndoGastric Solutions, Inc.•February 29, 2012

FDA WARNING_LETTER - EndoGastric Solutions, Inc. - February 29, 2012

View on Dashboard

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

ChatGPT

Claude

Perplexity

Grok

Copilot

Record Details

On June 8, 2012, the FDA issued a Warning Letter to Endogastric Solutions, Inc., following an inspection from February 10-29, 2012, regarding their clinical study, "A Randomized Controlled Trail of (b)(4) Patients to Reduce Regained Weight," IDE (b)(4), involving the (b)(4) investigational device. The inspection aimed to ensure compliance with federal regulations, data validity, and human subject protection.

The FDA identified several violations of 21 CFR Part 812 - Investigational Device Exemptions. Endogastric Solutions failed to provide investigators with necessary information, ensure proper monitoring, and promptly inform the IRB and FDA of significant new information (21 CFR 812.40). Specifically, monitoring visits at Site 01 and Site 02 were significantly delayed, exceeding protocol specifications. The firm also failed to promptly notify the IRB (notified April 2011) and FDA (not notified at all) when enrollment was put on hold in November/December 2010 due to questionable device effectiveness. Additionally, Endogastric Solutions did not provide for (b)(6) for clinical investigators.

Furthermore, Endogastric Solutions failed to notify FDA within 30 working days of the termination of the significant risk device investigation (21 CFR 812.150(b)(7)). The study was "cancelled

Company
EndoGastric Solutions, Inc.
Inspection Date
February 29, 2012
Product Type
Devices
Office
Center for Devices and Radiological Health
Person
  • Steven D. Silverman (President)
Open in Dashboard

ID · e37db5ef-49dd-46b1-8fcc-3b18ddbb63fd

Violation Codes6
21 U.S.C. 321(h)21 U.S.C. 360j(g)21 CFR 81221 CFR 812.4021 CFR 812.150(b)(7)21 CFR 812.3

Full citation text and observation details available on the Dashboard.

footer
Global Key Solutions
Regulatory Data Transformed
Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@globalkeysolutions.net
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026, Global Key Solutions All rights reserved.