FDA WARNING_LETTER - Hill-Rom Inc - October 11, 2011
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
On March 2, 2012, the FDA issued a Warning Letter to Hill-Rom, Inc. following an inspection from September 20 to October 11, 2011, at their Batesville, Indiana facility. The inspection found that the firm's AC-Powered Patient Lifts and AC-Powered Adjustable Hospital Beds were adulterated under 21 U.S.C. § 351(h) because their manufacturing, packing, storage, or installation methods did not conform to the Quality System (QS) regulation (21 CFR Part 820).
Key violations included: 1. **Design Verification (21 CFR 820.30(f)):** Failure to adequately verify device design, with design verification activities not confirming design output met input requirements, incomplete traceability summaries, and inadequate objective evidence for design requirement specifications. 2. **Corrective and Preventive Actions (CAPA) (21 CFR 820.100(a)):** Inadequate CAPA system, exemplified by 336 VersaCare bed battery failures without investigation or corrective action, and improper trending of discontinued components. 3. **Process Validation (21 CFR 820.75(a)):** Failure to fully validate automated/semi-automated processes for VersaCare beds, with an inadequate validation procedure only allowing periodic monitoring instead of full validation. 4
ID · 4c54f0ff-c285-427c-a357-2660fce27f69
Full citation text and observation details available on the Dashboard.