WARNING LETTER
•Hill-Rom, Inc.•June 21, 2007

FDA WARNING_LETTER - Hill-Rom, Inc. - June 21, 2007

View on Dashboard

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

ChatGPT

Claude

Perplexity

Grok

Copilot

Record Details

On January 15, 2008, the FDA issued a Warning Letter to Hill-Rom, Inc. following an inspection from May 29 to June 21, 2007, which found their VersaCare Beds (models P3200 and P3201) to be adulterated and misbranded.

The inspection revealed significant violations of Current Good Manufacturing Practice (CGMP) requirements under 21 CFR Part 820 (Quality System Regulation) and 21 CFR Parts 803 and 806 (Medical Device Reporting and Reports of Corrections and Removals).

Key violations include: 1. **Failure to establish and maintain corrective and preventive action (CAPA) procedures (21 CFR 820.100(a)):** Hill-Rom failed to fully implement CAPA00825, a design modification for an unacceptable latch failure rate. Despite redesigning the latching mechanism, only a portion of the affected 10,000 global units were corrected, with the firm initially classifying the issue as a "customer satisfaction issue" rather than a potential health hazard. While a proposed correction plan for all affected devices was later submitted, the initial response was deemed inadequate. 2. **Failure to establish and maintain adequate complaint handling procedures (21 CFR 820.198(c)):** The firm marked numerous complaints

Company
Hill-Rom, Inc.
Inspection Date
June 21, 2007
Product Type
Devices
Office
Detroit District Office
Person
  • Joanne M. Givens (Ms.)
Open in Dashboard

ID · a98ce998-8bac-40af-a55e-92ebafb082e3

Violation Codes10
21 U.S.C. 321(h)21 U.S.C. 351(h)21 CFR 82021 CFR 820.100(a)21 CFR 820.198(c)21 U.S.C. 352(t)(2)21 U.S.C. 360i21 CFR 80321 CFR 803.50(a)(2)21 CFR 806.10(a)(1)

Full citation text and observation details available on the Dashboard.

footer
Global Key Solutions
Regulatory Data Transformed
Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@globalkeysolutions.net
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026, Global Key Solutions All rights reserved.