FDA WARNING_LETTER - HoMedics, Incorporated - August 30, 2007
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
The FDA issued a Warning Letter to HoMedics, Inc. on June 16, 2008, following an inspection of their Dongguan, China facility from August 27-30, 2007. The letter identifies significant violations related to marketing clearance and Quality System (QS) regulation.
HoMedics is marketing the TheraP® Deluxe Automatic Blood Pressure Monitor and Therapist Select® Massagers (with infrared heating) in the U.S. without required marketing clearance or approval. The TheraP® monitor's claim "Irregular Heartbeat Detector" constitutes a new intended use, requiring a new 510(k) submission (21 CFR § 807.81(a)(3)), making the device misbranded (Section 502(o)) and adulterated (Section 501(f)(1)(B)). The Therapist Select® Massagers are also adulterated and misbranded for lacking marketing clearance, as infrared heating devices require a 510(k) (21 CFR § 890.5500). HoMedics is instructed to immediately cease promotional claims for the TheraP® monitor.
The inspection revealed adulterated heating pads, paraffin baths, and blood pressure monitors due to non-conformity with CGMP requirements of the QS regulation (21 CFR Part 820). Violations include: 1. Failure