WARNING LETTER
•Hoya Corporation - Pentax Life Division•April 24, 2015

FDA WARNING_LETTER - Hoya Corporation - Pentax Life Division - April 24, 2015

View on Dashboard

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

ChatGPT

Claude

Perplexity

Grok

Copilot

Record Details

The FDA issued a Warning Letter to Hoya Corporation (PENTAX Life Care Division) following inspections in April 2015 at their Japan facilities (Akishima-shi, Tokyo; Kurihara-shi, Miyagi) and Pentax of America, Inc. in Montvale, New Jersey. The inspections found that the firm's endoscopes and accessories were adulterated under 21 U.S.C. § 351(h) due to non-conformity with Quality System (QS) regulation (21 CFR Part 820) and misbranded under 21 U.S.C. § 352(t)(2) for failing to furnish required information under 21 U.S.C. § 360i and 21 CFR Part 803 (Medical Device Reporting - MDR).

Key violations include: 1. **Failure to establish and maintain design validation procedures (21 CFR 820.30(g))**: Validation studies for ED-3670TK's EtO sterilization and cleaning/HLD IFUs used different model/series endoscopes without documented justification for applicability. EtO sterilization validation for ED-3490TK and ED-3670TK used incorrect gas concentrations. DHF lacked protocol and raw data for EtO sterilization and HLD validation reports. Reprocessing validation protocol lacked specific testing conditions (

Company
Hoya Corporation - Pentax Life Division
Inspection Date
April 24, 2015
Product Type
Devices
Office
Center for Devices and Radiological Health
People
  • Jan B. Welch
  • Diana Amador-Toro (District Director)
  • Ganesh Ramaswamy
  • Christopher Burton
  • Hiroshi Suzuki
Open in Dashboard

ID · 3eccb8e9-22b4-45b8-9a2a-c2a2ce5bbc91

Violation Codes10
21 U.S.C. 321(h)21 U.S.C. 351(h)21 CFR 82021 CFR 820.30(g)21 CFR 820.30(f)21 CFR 820.100(a)21 CFR 820.75(a)21 U.S.C. 352(t)(2)21 U.S.C. 360i21 CFR 803

Full citation text and observation details available on the Dashboard.

footer
Global Key Solutions
Regulatory Data Transformed
Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@globalkeysolutions.net
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026, Global Key Solutions All rights reserved.