FDA WARNING_LETTER - Hoya Corporation - Pentax Life Division - April 24, 2015
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
The FDA issued a Warning Letter to Hoya Corporation (PENTAX Life Care Division) following inspections in April 2015 at their Japan facilities (Akishima-shi, Tokyo; Kurihara-shi, Miyagi) and Pentax of America, Inc. in Montvale, New Jersey. The inspections found that the firm's endoscopes and accessories were adulterated under 21 U.S.C. § 351(h) due to non-conformity with Quality System (QS) regulation (21 CFR Part 820) and misbranded under 21 U.S.C. § 352(t)(2) for failing to furnish required information under 21 U.S.C. § 360i and 21 CFR Part 803 (Medical Device Reporting - MDR).
Key violations include: 1. **Failure to establish and maintain design validation procedures (21 CFR 820.30(g))**: Validation studies for ED-3670TK's EtO sterilization and cleaning/HLD IFUs used different model/series endoscopes without documented justification for applicability. EtO sterilization validation for ED-3490TK and ED-3670TK used incorrect gas concentrations. DHF lacked protocol and raw data for EtO sterilization and HLD validation reports. Reprocessing validation protocol lacked specific testing conditions (