FDA WARNING_LETTER - Jieying Laboratory Inc - October 30, 2014
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
On March 9, 2015, the FDA issued a Warning Letter to Jieying Laboratory, Inc. following an inspection from October 27-30, 2014, in Longueuil, Quebec, Canada. The inspection revealed that the firm's Class II assisted reproduction tools are adulterated under 21 U.S.C. § 351(h) because their manufacturing, packing, storage, or installation methods do not conform to the Quality System (QS) regulation (21 CFR Part 820).
The letter details numerous violations, including: 1. **Failure to establish and maintain design control procedures** (21 CFR 820.30(a)), lacking requirements for design planning, input, output, review, verification, validation, transfer, changes, and design history files. 2. **Failure to validate processes** where results cannot be fully verified by subsequent inspection and test (21 CFR 820.75(a)), specifically for sterile micro-tool manufacturing and sterilization processes (e.g., lack of yearly bioburden/sterility recovery testing, revalidation for material changes). 3. **Failure to establish procedures for nonconforming product review and disposition** (21 CFR 820.90(b)(1)). 4. **Failure to develop, conduct, control, and monitor production processes** to ensure device conformity (2
ID · 41204700-ab75-4eea-adb6-7f01c4b5cd2d
Full citation text and observation details available on the Dashboard.