Record Details

On August 11, 2014, the FDA issued a Warning Letter to Marel Corporation following an inspection from May 30 to June 24, 2014. The inspection determined Marel Corporation is a repackager/relabeler of sterile towel/drapes, classified as devices under section 201(h) of the Act. The devices were found to be adulterated under section 501(h) of the Act, as manufacturing, packing, storage, or installation methods did not conform to the Quality System Regulation (21 CFR Part 820).

Violations included: 1. Failure to validate processes (21 CFR 820.75(a)): No validation documentation for ETO sterilization's effect on packaging, or post-sterilization package integrity testing. Pre-sterilization packaging validation was inadequate, lacking specific packaging/machinery details and material identification. The "Packaging Machine Validation" procedure was incomplete, missing a page from the test method and acceptance criteria for seals. 2. Failure to monitor and control process parameters for validated processes (21 CFR 820.75(b)): No requalification assessments for the ETO sterilization process since 2010, despite a requirement for (b)(4) assessments. 3. Failure to control and monitor production processes (21 CFR 820.70(

Company
Marel Corporation
Inspection Date
June 24, 2014
Product Type
Devices
Office
New England District Office
Person
  • Mutahar Shamsi (District Director)
Open in Dashboard

ID · ce592ba8-5dd4-4662-b7e2-eb9855f41095

Violation Codes6
21 U.S.C. 321(h)21 U.S.C. 351(h)21 CFR 82021 CFR 820.75(a)21 CFR 820.75(b)21 CFR 820.70(a)

Full citation text and observation details available on the Dashboard.

footer
Global Key Solutions
Regulatory Data Transformed
Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@globalkeysolutions.net
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026, Global Key Solutions All rights reserved.