FDA WARNING_LETTER - Medical Solutions International, Inc. - March 17, 2008
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
On March 5-17, 2008, an FDA inspection of Medical Solutions International, Inc. in Merriam, Kansas, revealed that their BiPro SDS-CB200 Bicarb Mixing and Distribution Systems, classified as devices, are adulterated. This is due to non-conformity with Current Good Manufacturing Practice (CGMP) requirements of the Quality System (QS) regulation (21 CFR Part 820).
Key violations include: - **Failure to establish and maintain document control (21 CFR 820.40):** No document control procedures, unapproved/obsolete documents in use, and unapproved document changes. - **Failure to establish and maintain written procedures in multiple areas:** - Identifying training needs (21 CFR 820.25(b)). - Identifying product throughout all stages (21 CFR 820.60). - Changes to specifications, processes, and procedures (21 CFR 820.70(b)). - Labeling (21 CFR 820.120). - Preventing mix-ups, damage, or adverse effects during handling (21 CFR 820.140). - Controlling product storage (21 CFR 820.150(a)). - Controlling and distributing finished devices (21 CFR 8