FDA WARNING_LETTER - MicroVention Costa Rica S.R.L. - May 26, 2022
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
On September 30, 2022, the FDA issued a Warning Letter to Charlie Noel following an inspection from May 23-26, 2022, at their Alajuela, Costa Rica facility. The inspection revealed that the firm's medical devices, including Headway Microcatheter, LVIS, LVIS Jr, Scepter C and XC Occlusion Balloon Catheter, SOFIA Catheters, and BOBBY Balloon Guide Catheter, were adulterated under 21 U.S.C. § 351(h) due to non-conformity with Quality System (QS) regulation (21 CFR Part 820).
Key violations included: 1. **Failure to verify processes (21 CFR 820.75(a))**: The catheter wash limitation process validation lacked specified requirements, and justifications did not provide high assurance or approval. Biocompatibility testing (ISO 10993-1) was inadequate, missing pyrogenicity assessment, incorrect cytotoxicity acceptance criteria, and failure to address residual sterilant levels (ISO 10993-7). Software validation for the coating tracker lacked evidence of required repeat tests. The firm's CAPA (b)(4) was deemed inadequate. 2. **Inadequate CAPA procedures (21 CFR 820.100)**: The Supplier Corrective Action Request (SC