WARNING LETTER
•Middlesboro Appalachian Regional Hospital

FDA WARNING_LETTER - Middlesboro Appalachian Regional Hospital

View on Dashboard

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

ChatGPT

Claude

Perplexity

Grok

Copilot

Record Details

On August 18, 2011, an inspection by a representative of the Commonwealth of Kentucky, acting on behalf of the FDA, revealed serious violations of the Mammography Quality Standards Act of 1992 (MQSA) at Middlesboro Appalachian Regional Hospital. The inspection identified two primary deficiencies. First, the facility failed to provide documentation verifying that two radiologic technologists met the continuing experience requirement of performing 200 mammography examinations in 24 months, a repeat Level 2 violation of 21 CFR 900.12(a)(2)(iv). Second, a mammography equipment evaluation by a medical physicist for a specific laser printer was not conducted, a Level 2 violation of 21 CFR 900.12(e)(10). The facility's response on August 31, 2011, was deemed inadequate as it lacked a plan for permanent corrective action. This repeat violation regarding technologist experience has been cited in multiple previous annual MQSA inspections dating back to 2001, indicating serious underlying problems that could compromise mammography quality. The FDA warns that continued non-compliance may lead to additional actions, including an Additional Mammography Review, a Directed Plan of Correction, civil money penalties up to $10,000 per failure, suspension or revocation of the FDA certificate, and fee-based compliance follow-up inspections. The facility is required to respond within fifteen working days, detailing specific corrective and preventative steps, including timeframes, and providing sample records.

Company
Middlesboro Appalachian Regional Hospital
Product Type
Devices
Office
Cincinnati District Office
Person
  • Cheryl A. Bigham (Division Director)
Open in Dashboard

ID · fcf106d9-59b8-452a-89fb-0dea8d1512b9

Violation Codes4
21 CFR 900.12(a)(2)(iv)21 CFR 900.12(e)(10)21 CFR 900.12(j)42 U.S.C. 263b(h)-(j)

Full citation text and observation details available on the Dashboard.

footer
Global Key Solutions
Regulatory Data Transformed
Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@globalkeysolutions.net
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026, Global Key Solutions All rights reserved.