FDA WARNING_LETTER - NeuroSync, Inc. - February 05, 2025
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
On May 9, 2025, the FDA issued a Warning Letter to a firm in Holliston, MA, following an inspection from January 6 to February 5, 2025. The firm manufactures EYE-SYNC, a virtual reality headset and tablet for eye assessments, classified as a medical device. The inspection revealed the devices are adulterated under 21 U.S.C. § 351(h) due to non-conformity with Quality System (QS) regulation (21 CFR Part 820).
Key violations include: 1. **Failure to establish and maintain design control procedures (21 CFR 820.30(a))**: * Failure to verify device design against input requirements (21 CFR 820.30(f)), e.g., EYE-SYNC accuracy requirement not met, and lack of testing for data quality warnings. * Failure to identify, document, validate, verify, review, and approve design changes before implementation (21 CFR 820.30(i)), e.g., software correction not verified. * Failure to establish appropriate design inputs (21 CFR 820.30(c)), e.g., outdated requirements in design input documents. * Failure to validate device design to user needs (21 CFR 820.30
ID · 407f9de3-8676-477c-b52a-7f678fd90b43