FDA WARNING_LETTER - Northern Digital Inc. - March 28, 2013
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
On July 24, 2013, the FDA issued a Warning Letter to Northern Digital, Inc. (NDI) following a March 25-28, 2013, inspection of their Waterloo, ON facility, which manufactures Reflective Passive Marker Spheres for Image Guided Surgical Systems. The inspection found these devices adulterated under 21 U.S.C. § 351(h) due to non-conformity with Quality System (QS) regulation (21 CFR Part 820) and misbranded under 21 U.S.C. § 352(t)(2) for failing to provide required information under 21 U.S.C. § 360i and 21 CFR Part 803 (Medical Device Reporting - MDR).
Key violations include: 1. **Process Validation (21 CFR 820.75(a)):** Failure to validate the sphere manufacturing process, especially after relocation to Mexico in 2009. IQ/OQ/PQ was not performed during the relocation or prior manufacturing in Waterloo. Corrective actions are ongoing, including a retrospective review, CAPA CAR157, and audits of the contract manufacturer, with a target completion by August 30, 2013. 2. **Design Verification (21 CFR 820.30
ID · 3aafa020-8203-4059-b952-ff85f50e3018
Full citation text and observation details available on the Dashboard.