On August 5-8, 2013, an FDA inspection of Rehabtronics, Inc. in Edmonton, Alberta, Canada, revealed that their Saebo ReJoyce upper-extremity rehabilitation device was adulterated under 21 U.S.C. § 351(h) due to non-conformity with Quality System (QS) regulation 21 CFR Part 820. The firm's August 28, 2013 response to the FDA 483 was deemed inadequate.

Violations include: 1. **Failure to establish and maintain CAPA procedures (21 CFR 820.100(a)):** The CAPA procedure lacked requirements for analyzing quality data using statistical methodology to identify nonconforming product causes. The firm's proposed changes did not explain how data would be used or how statistical methods would detect recurring problems. 2. **Failure to maintain complaint files and procedures (21 CFR 820.198(a)):** (b)(4) complaints from 2011-2013 lacked Medical Device Report (MDR) evaluations. The firm failed to conduct a retrospective review for MDR reportability and did not provide revised procedures or evidence of implementation. 3. **Failure to investigate complaints (21 CFR 820.198(c)):** (b)(4) complaints

Company
Rehabtronics Inc.
Inspection Date
August 8, 2013
Product Type
Devices
Office
Center for Devices and Radiological Health
Person
  • Steven D. Silverman (President)
Open in Dashboard

ID · b7eb0d43-655e-42be-9d6e-2e3d92c6032a

Violation Codes9
21 U.S.C. 321(h)21 U.S.C. 351(h)21 CFR 82021 CFR 820.100(a)21 CFR 820.198(a)21 CFR 820.198(c)21 CFR 820.30(g)21 CFR 820.18421 CFR 820.20(c)

Full citation text and observation details available on the Dashboard.

footer
Global Key Solutions
Regulatory Data Transformed
Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@globalkeysolutions.net
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026, Global Key Solutions All rights reserved.