FDA WARNING_LETTER - Simply Clean Air & Water Inc. - July 18, 2014
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
On August 18, 2014, the FDA issued a Warning Letter to Simply Clean Air & Water Inc. following an inspection from July 9-18, 2014. The inspection revealed that the firm's Portable Exchange Deionization (PEDI) systems, classified as medical devices, were adulterated under 21 U.S.C. § 351(h) due to non-conformity with Quality System (QS) regulation (21 CFR Part 820).
Key violations included: 1. **Failure to establish and maintain design validation procedures (21 CFR 820.30(g))**: PEDI systems for hemodialysis were labeled to an outdated standard (ANSI/AAMI RD62:2001), lacking documentation of testing to the current standard or any design control procedures. 2. **Failure to establish and maintain procedures for acceptance activities (21 CFR 820.80(a))**: No inspection, test, or acceptance criteria existed for manufacturing materials, with incoming components placed directly into inventory. 3. **Failure to establish and maintain procedures for purchased products (21 CFR 820.50)**: The firm lacked supplier qualification forms and records of inspection/testing for incoming materials. 4. **Failure to maintain a device history record (DHR) (21 CFR 82