FDA WARNING_LETTER - SMT-Schilling Metalltechnik GmbH - July 19, 2012
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
On October 9, 2012, the FDA issued a Warning Letter to SMT-Schilling Metalltechnik GmbH following a July 16-19, 2012 inspection in Muhlheim, Germany. The inspection found that the firm's Class II K/Guide Wires were adulterated under 21 U.S.C. § 351(h) due to non-conformity with Quality System (QS) regulation (21 CFR Part 820) and misbranded under 21 U.S.C. § 352(t)(2) for failing to furnish required information under 21 U.S.C. § 360i and 21 CFR Part 803 (Medical Device Reporting - MDR).
Key violations included: 1. **CAPA (21 CFR 820.100(a)):** Failure to establish and maintain adequate CAPA procedures, including investigation of non-conformances and analysis of quality data. The firm's response was deemed adequate for this point. 2. **Design Control (21 CFR 820.30(a)):** No procedures or documentation for design control (planning, inputs, outputs, reviews, verification, validation, transfer) for the K-Wire/Guide Wire design project. Response was inadequate. 3. **Purchasing Controls
ID · 041244eb-f552-43f1-98cc-334ce271f758