WARNING LETTER
•Sonesta Medical AB•June 7, 2023

FDA WARNING_LETTER - Sonesta Medical AB - June 07, 2023

View on Dashboard

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

ChatGPT

Claude

Perplexity

Grok

Copilot

Record Details

The FDA issued a Warning Letter to Industrivagen 7, Solna, Sweden, following an inspection from June 5-7, 2023. The firm manufactures medical devices, specifically the Model 6210 radiographic table and Model S2 and S3 cystometric tables. These devices were found to be adulterated under section 501(h) of the Act, as their manufacturing, packing, storage, or installation methods do not conform to the Quality System (QS) regulation (21 CFR Part 820).

Key violations include: 1. **Purchasing Controls (21 CFR 820.50(a)):** Failure to establish and maintain procedures for purchased products and services to conform to requirements. The firm's Supplier Evaluation procedure lacks defined frequency, type, and extent of control over suppliers, and the contract repair/servicer and QA/RA consultant were not evaluated. The firm's responses were inadequate, as assessments are incomplete, and revised procedures were not provided. 2. **Complaint Handling (21 CFR 820.198(a)):** Failure to establish and maintain procedures for receiving, reviewing, and evaluating complaints. The "Process description deviation management" procedure lacks requirements for Medical Device Reporting evaluation. Complaint records inconsistently include investigation dates/results and often omit Unique Device Identification (UDI) codes. Responses were inadequate, with incomplete reviews and unprovided revised

Company
Sonesta Medical AB
Inspection Date
June 7, 2023
Product Type
Devices
Office
Center for Devices and Radiological Health
People
  • Greg Swanson
  • Robert A. Ochs (Office Director)
Open in Dashboard

ID · e64818d2-e29e-4938-accf-a97fc8ec5b29

Violation Codes9
21 U.S.C. 321(h)21 U.S.C. 351(h)21 CFR 82021 CFR 820.50(a)21 CFR 820.198(a)21 CFR 820.30(c)21 CFR 820.30(g)21 CFR 820.30(i)21 CFR 820.40

Full citation text and observation details available on the Dashboard.

footer
Global Key Solutions
Regulatory Data Transformed
Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@globalkeysolutions.net
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026, Global Key Solutions All rights reserved.