FDA WARNING_LETTER - Spacelabs Healthcare Inc - April 02, 2014
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
This FDA Warning Letter, dated August 11, 2014, is an amendment to an August 5, 2014 letter, correcting a typographical error and extending the response deadline. It resulted from an inspection of a Snoqualmie, Washington firm manufacturing patient monitoring devices (e.g., Qube, Ultraview SL, ?lance Central Station) from February 12 to April 2, 2014. The inspection found the devices adulterated under 21 U.S.C. § 351(h) due to non-conformity with Quality System (QS) regulation (21 CFR Part 820).
Violations include: 1. **Failure to establish and maintain CAPA procedures (21 CFR 820.100(a))**: CAPAs were closed without implementing corrective actions (e.g., CAPA 31920, 24340) and effectiveness was not verified/validated (e.g., CAPA 31363). 2. **Failure to control nonconforming product (21 CFR 820.90(a))**: Out-of-specification components were accepted without Nonconforming Materials Records (NMRs). 3. **Failure to establish and maintain complaint handling procedures (21 CFR 820.198(a))**: Service
ID · ea469af3-62b0-4e71-9927-8ffef17d2531