FDA WARNING_LETTER - Spinetronics, LLC - March 01, 2011
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
On February 28 - March 1, 2011, an FDA inspection of Spinetronics, LLC in Coral Springs, Florida, revealed that their Antalgic-Trak device was adulterated and misbranded. The device was adulterated under 21 U.S.C. § 351(h) due to non-conformity with Quality System (QS) regulation (21 CFR Part 820) and under 21 U.S.C. § 351(f)(1)(B) for lacking marketing clearance or approval.
Violations of QS regulation included: - Failure to establish and maintain design control procedures (21 CFR 820.30(a)(1)), lacking documentation for design validation, design change controls, risk analysis, or structural testing of embedded software. - Failure to establish and maintain procedures for finished device acceptance (21 CFR 820.80(d)). - Failure to establish and maintain purchasing control procedures (21 CFR 820.50), including lack of awareness of contract manufacturer's quality system activities. - Failure to establish and maintain nonconforming product procedures (21 CFR 820.90(a)). - Failure to establish and maintain Corrective and Preventive Action (CAPA) procedures (21 CFR 820.100(a)). - Failure