FDA WARNING_LETTER - The Lebanon Corp., Inc. - February 24, 2012
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
The Lebanon Corp., Inc. received an FDA Warning Letter (2012-DET-13) on May 17, 2012, following an inspection from December 19, 2011, to February 24, 2012. The inspection revealed that the firm's Honan Intraocular Pressure Reducer device was adulterated under 21 U.S.C. § 351(h) due to non-conformity with Quality System (QS) regulation (21 CFR Part 820).
Key violations included: 1. **Failure to establish and maintain adequate CAPA procedures (21 CFR 820.100(a))**: The firm's procedure lacked details on quality data analysis means and frequency, despite numerous receiving, in-process, and final inspection failures (1045, 661, and 12 respectively from 2007-2012). 2. **Failure to establish and maintain adequate complaint handling procedures (21 CFR 820.198(a))**: Procedures did not define a complaint, specify required documentation, outline follow-up activities, or define evaluation criteria for reportable events. The firm claimed no complaints, but 32 uninvestigated service reports were found. 3. **Failure to control non-conforming product (2
ID · 7bc2f819-9528-4668-8946-7df824802056