FDA WARNING_LETTER - Unetixs Vascular, Inc. - October 29, 2019
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
The FDA issued a Warning Letter to Unetixs Vascular Inc. on February 4, 2020, following an inspection from September 17 to October 29, 2019. The inspection revealed that Unetixs, a manufacturer of vascular diagnostic ultrasound systems like the MultiLab® Series, had adulterated devices under 21 U.S.C. § 351(h) due to non-conformity with current good manufacturing practice requirements of the Quality System regulation (21 CFR Part 820).
Significant violations include: 1. **Failure to control device design (21 CFR § 820.30(a))**: Specifically, the MultiLab® Series ROODRA, distributed since September 28, 2018, lacked established design and development plans, design inputs, design outputs, design reviews, and design validation. A Design History File (DHF) compliant with FDA requirements was not established prior to distribution. 2. **Failure to control production processes (21 CFR § 820.70(a))**: No approved procedures existed for ROODRA manufacturing, including testing for electrical leakage, acoustic output, EMC, and electrical safety. 3. **Failure to establish acceptance activities procedures (21 CFR § 820.80(a))**: Four lots of Revo motherboards, a critical component
ID · a65f7ed2-4a62-4333-a2a8-6226d0a719f8
Full citation text and observation details available on the Dashboard.