FDA WARNING_LETTER - WEBER MEDICAL GMBH - January 17, 2013
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
On January 14-17, 2013, an FDA inspection of Weber Medical GmbH in Lauenfoerde, Germany, revealed that their Weberneedle device is adulterated under 21 U.S.C. § 351(h) because its manufacturing, packing, storage, or installation methods, facilities, or controls do not conform to the Quality System (QS) regulation (21 CFR Part 820).
Violations include: - **Failure to establish and maintain design validation procedures (21 CFR 820.30(g)):** The user manual lacks instructions for the patient switch, and the firm's response lacked documentation of corrections. - **Failure to evaluate and select suppliers based on quality requirements (21 CFR 820.50(a)(1)):** Supplier (b)(4) lacks a quality management system and their contract doesn't include FDA QS regulation requirements. The firm's response lacked documentation of corrections. - **Failure to define control over suppliers based on evaluation results (21 CFR 820.50(a)(2)):** The supplier grading system doesn't specify criteria for required corrective actions. The firm provided no written response to this observation. - **Failure to maintain Device History Records (DHRs) (21 CFR 820.184):** No DHRs exist for components manufactured by