FDA WARNING_LETTER - Xoran Technologies Inc - October 26, 2007
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
On March 25, 2008, the FDA issued a Warning Letter to Xoran Technologies, Inc. following an inspection from August 30, 2007, to October 26, 2007. The inspection found that the firm's MiniCAT CT System, a medical device, was adulterated under section 501(h) of the Federal Food, Drug, and Cosmetic Act, due to non-conformity with Current Good Manufacturing Practice (CGMP) requirements of the Quality System regulation (21 CFR Part 820).
Violations included: 1. Failure to document design validation results in Design History Files (21 CFR § 820.30(g)). 2. Failure to maintain a complete Device Master Record (21 CFR § 820.181). 3. Failure to document proper installation procedures and results (21 CFR § 820.170(b)). 4. Failure to maintain Device History Records with primary identification labels (21 CFR § 820.184(e)). 5. Failure to control document approval and distribution (21 CFR § 820.40(a)). 6. Failure to validate and document computer software used in production or the quality system (21 CFR § 820.70(i)).
ID · 30c3deea-2685-44a2-a19e-d7f28540e823
Full citation text and observation details available on the Dashboard.