483
•Terumo Cardiovascular Systems Corporation•May 18, 2004

Terumo Cardiovascular Systems Corporation – FDA 483, May 18, 2004

Record Details

Terumo Cardiovascular Systems Corp., a medical device manufacturer in Ann Arbor, Michigan, received an FDA Form 483 following an inspection conducted from May 10 to May 18, 2004. The document outlines significant observations regarding the company's adherence to medical device quality system regulations.

Key issues identified span several critical areas. In Management Controls, the FDA noted an inadequate and ineffective quality system, including insufficient quality audit procedures, lack of regular audits, and employees unfamiliar with the company's quality policy. Design Controls revealed unresolved discrepancies post-design validation, incomplete risk analyses, and unapproved design plans.

Corrective and Preventive Actions (CAPA) were cited for incomplete procedures, failure to assess complaints for Medical Device Report (MDR) reportability, untimely processing of numerous complaints, and lack of verification for corrective actions. Furthermore, Production and Process Controls had deficiencies such as unvalidated processes, improper document control for changes, uninvestigated environmental control excursions, uncalibrated equipment, delayed employee training, and inadequate product storage controls.

The FDA expects Terumo Cardiovascular Systems Corp. to respond to these observations with detailed corrective actions and a plan for their implementation to ensure compliance with federal regulations governing medical device manufacturing.

Company
Terumo Cardiovascular Systems Corporation
Inspection date
May 18, 2004
Product type
Devices
Office
Detroit District Office
People
  • Kinh Quoc Mac
  • Art O. Czabaniuk (Program Division Director)
View Company Profile

ID · a921663c-9a8e-413a-aab8-c4f5bd9eab7d

Violation Codes2
21 CFR 21021 CFR 211

Full citation text and observation details available on the Dashboard.

Other documents for Terumo Cardiovascular Systems Corporation
  • 483 - 2012-09-28September 28, 2012
  • 483 - 2010-03-29March 29, 2010
  • 483 - 2008-08-11August 11, 2008
  • 483 - 2005-12-09December 9, 2005

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore Terumo Cardiovascular Systems Corporation
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.