Medtronic MiniMed, Inc. – FDA EIR, Mar 14, 2018

Record Details

An FDA inspection of Medtronic MiniMed, Inc. in Northridge, CA, was conducted from February 14 to March 14, 2018. The inspection focused on postmarket activities for the MiniMed 670G Pump and Guardian Sensor (3), operating under regulatory frameworks for medical device manufacturers and PMA postmarket inspections (Compliance Programs 7383.001 and 7382.845). The review covered design controls, corrective and preventive actions (CAPA), and production and process controls.

Two significant issues were identified and documented on an FDA-483. First, Medtronic MiniMed failed to submit 4,458 Medical Device Reports (MDRs) within the required 30-day period in 2017 for device malfunctions likely to cause or contribute to death or serious injury if they recurred (21 CFR 803.50(a)(2)). Second, the company did not adequately review, evaluate, or investigate complaints, particularly those leading to MDRs involving serious adverse events or injuries related to the MiniMed 670G Pump, failing to prioritize these critical investigations in a timely manner (21 CFR 820.198(c)).

Company management agreed with both observations, promising corrective measures. They committed to submitting a written response within 15 business days and discussed improvements to CAPA timelines, non-conformity investigation documentation, and trend analysis.

Company
Medtronic MiniMed, Inc.
Inspection date
March 14, 2018
Product type
devices
Office
Los Angeles District Office
Person
  • Sarah A. Hassas (investigator)
View Company Profile

ID · 79bfab82-c285-493a-9d50-a17c4839c5a0

Violation Codes8
21 CFR 820.19821 CFR 820.3021 CFR 820.10021 CFR 820.198(c)21 CFR 80321 CFR 803.50(a)(2)21 CFR 820.5021 CFR 800-1299

Full citation text and observation details available on the Dashboard.

Other documents for Medtronic MiniMed, Inc.
  • EIR - 2023-03-10March 10, 2023
  • 483 - 2023-03-10March 10, 2023
  • 483 - 2021-07-07July 7, 2021
  • EIR - 2019-02-15February 15, 2019
  • 483 - 2018-03-14March 14, 2018

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore Medtronic MiniMed, Inc.
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.