FDA Inspection
Astura MedicalAstura Medical – FDA Inspection, Apr 23, 2026
Record Details
This FDA Inspection record concerns Astura Medical, with an inspection on April 23, 2026, issued by the Dallas District Office, covering medical devices & rad health.
View Company Profile
ID · d0973835-9543-4d16-9d15-ae0b2d5be028
Inspection Facts
Voluntary Action Indicated (VAI)
Objectionable conditions were found, but FDA is not prepared to take or recommend regulatory action.
- Inspection number
- 1313620
- FDA center
- Medical Devices & Rad Health
- Project area
- Compliance: Devices
FDA Citations
5- ISO 13485:2016 Clause 7.3.9Procedures to control design and development changes not documented
- ISO 13485:2016 Clause 7.4.1Supplier performance not monitored
- ISO 13485:2016 Clause 7.5.6Process for production and service provision not validated
- ISO 13485:2016 Clause 8.2.1Information gathered in feedback process not used as input into risk management
- ISO 13485:2016 Clause 8.2.2Complaint handling procedures lack required elements
Full citation text and observation details available on the Dashboard.