WARNING LETTER
•Ascension Orthopedics, Inc•March 8, 2010

Ascension Orthopedics, Inc – FDA Warning Letter, Mar 8, 2010

Record Details

The FDA issued a Warning Letter to Ascension Orthopedics, Incorporated following an inspection from March 3-8, 2010, concerning their Interphalangeal (PIP) Trial devices. The inspection revealed that the devices were misbranded due to the firm's failure to furnish required information under sections 502(t)(2) and 519 of the Act, specifically regarding Reports of Corrections and Removals (21 CFR Part 806) and Medical Device Reporting (MDR) (21 CFR Part 803).

Key violations included the failure to submit a written report to FDA within 10 working days for a Class II recall of oversized trials (21 CFR 806.10(a)(1)). The firm also failed to submit MDRs within 30 days for serious injuries (e.g., bone fracture requiring suture) (21 CFR 803.50(a)(1)) and for device malfunctions that would likely cause serious injury if they recurred (21 CFR 803.50(a)(2)). Furthermore, Ascension Orthopedics failed to develop, implement, and maintain adequate written MDR procedures (21 CFR 803.17). An additional nonconformity with Quality System (QS) regulation (21 CFR Part 820) was the failure to properly review, evaluate, and investigate complaints, as evidenced by incorrect conclusions regarding patient harm in a complaint involving a bone fracture (21 CFR 820.198(c)).

The FDA deemed the firm's responses incomplete or inadequate, noting a lack of documented corrective actions and updated procedures. Ascension Orthopedics must take prompt action to correct these violations within fifteen working days, providing documentation and a timetable for implementation. Failure to comply may result in regulatory actions such as seizure, injunction, civil money penalties, and impact on federal contracts.

Company
Ascension Orthopedics, Inc
Inspection date
March 8, 2010
Product type
Devices
Office
Dallas District Office
Person
  • Reynaldo Rodriguez (District Director)
View Company Profile

ID · 83e637ee-2be2-4d70-be22-267171f337d8

Violation Codes10
21 U.S.C. 321(h)21 U.S.C. 352(t)(2)21 U.S.C. 360i21 CFR 80621 CFR 806.10(a)(1)21 CFR 80321 CFR 803.50(a)(1)21 CFR 803.50(a)(2)21 CFR 803.1721 CFR 820.198(c)

Full citation text and observation details available on the Dashboard.

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore Ascension Orthopedics, Inc
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.