Record Details

On April 15-19, 2013, an FDA inspection of Electro Medical Inc. in Tulsa, Oklahoma, revealed that the firm's Stim Flex 400, 400A, 150, and 500 devices are adulterated under 21 U.S.C. § 351(h) due to non-conformity with Quality System (QS) regulation (21 CFR Part 820). The firm's April 30, 2013, response to the FDA 483 was deemed inadequate.

Violations include: 1. **Failure to establish design control procedures (21 CFR 820.30(a))**: No documented design inputs, outputs, verification, validation, changes, or history files for Stim Flex 150 and 500. Response lacked corrective actions for missing documentation and failed to adequately address risk analysis for Stim Flex 150's technological differences. 2. **Inadequate acceptance activity procedures (21 CFR 820.80(a))**: In-process inspection records were discarded, and no acceptance records for 7 Stim Flex 400A and approximately 40 Stim Flex 150 units were available. Response lacked timeframe and specific corrective actions. 3. **Failure to perform quality audits (21 CFR

Company
Electro Medical, Inc.
Inspection date
April 19, 2013
Product type
Devices
Office
Dallas District Office
Person
  • Reynaldo Rodriguez (District Director)
View Company Profile

ID · d70ff256-d490-44be-8b71-65db2d1751e7

Violation Codes10
21 U.S.C. 321(h)21 U.S.C. 351(h)21 CFR 82021 CFR 820.30(a)21 CFR 820.80(a)21 CFR 820.2221 CFR 820.19821 CFR 820.18121 CFR 820.18421 U.S.C. 351(f)(1)(B)

Full citation text and observation details available on the Dashboard.

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore Electro Medical, Inc.
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.