WARNING LETTER
•Instratek, Incorporated•April 24, 2012

Instratek, Incorporated – FDA Warning Letter, Apr 24, 2012

Record Details

During an April 2-24, 2012, inspection, the FDA found Instatek, Inc., a manufacturer of orthopedic implant systems (e.g., Sub-Talar Lok, Michelangelo Bunion System), in violation of Quality System (QS) regulation (21 CFR Part 820) requirements, rendering their devices adulterated. The firm's responses from May to July 2012 were incomplete or partially adequate.

Violations included: 1. **Design Validation (21 CFR § 820.30(g)):** Failure to validate pre-vacuum steam sterilization cycles for Sub-Talar Lok and Michelangelo Bunion System (MABS) and to adequately address gravity displacement steam sterilization. Risk analysis for Sub-Talar Lok instruments was incomplete, and a design validation plan/protocol was missing. 2. **Design Input (21 CFR § 820.30(c)):** Incomplete or ambiguous design input requirements for the Sub-Talar Lok Implant System and its accessories, lacking technical specifications and measurable requirements. 3. **Finished Device Acceptance (21 CFR § 820.80(d)):** Incorrect measurement of Sub-Talar implants during inspection and failure to document nonconforming inspection results for rejected lots. 4. **Corrective and Preventive Action (CAPA) (21 CFR § 820.10

Company
Instratek, Incorporated
Inspection date
April 24, 2012
Product type
Devices
Office
Dallas District Office
Person
  • Reynaldo Rodriguez (District Director)
View Company Profile

ID · 9e1efc21-629a-4f77-a58e-249b6062c478

Violation Codes8
21 U.S.C. 321(h)21 U.S.C. 351(h)21 CFR 82021 CFR 820.30(g)21 CFR 820.30(c)21 CFR 820.80(d)21 CFR 820.10021 CFR 820.198(a)

Full citation text and observation details available on the Dashboard.

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

KeyPedia

ChatGPT

Claude

Perplexity

Grok

Copilot

Document or profile

One record shows the event. A profile shows the pattern.

Buy this document for the source file and details behind this single event.

Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.

Explore Instratek, Incorporated
KeyPedia
Regulatory Data Transformed
Stay in the loop

KRA score changes, company updates and news. About monthly, unsubscribe anytime.

Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@keypedia.com
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026 KeyPedia™ | Global Key Solutions Corp. All rights reserved.