483
GRAMERCY EXTREMITY ORTHOPEDICSFDA 483 - GRAMERCY EXTREMITY ORTHOPEDICS - October 24, 2025
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
Record Details
GRAMERCY EXTREMITY ORTHOPEDICS in Richardson, TX, a manufacturer of class I and II medical devices, was cited for failing to adequately establish process control procedures. Specifically, the firm did not include titanium plate implants in their sterilization procedure's dose audit requirements and failed to conduct these audits as specified before distributing finished devices. This indicates a significant lapse in ensuring product conformance to specifications.
- Office
- Dallas District Office
- Person
Open in Dashboard
ID · 420690d7-1865-4be8-84b0-fe6b7652cf55