FDA 483 - Pacific Consolidated Industries LLC - July 25, 2025
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
An FDA inspection of Pacific Consolidated Industries LLC in Riverside, CA, a medical device manufacturer, revealed significant deficiencies in their quality system. The firm's procedures for corrective and preventive actions (CAPA) were found to be inadequately established and implemented, with issues in classification, timely completion, and record-keeping. Additionally, procedures for design change were insufficient, lacking proper justification, risk assessments, verification, and validation for engineering changes, and inconsistent use of a 510(k) decision flow chart.
- Person
ID · 06922644-e6da-4774-aa43-a8f28a8a8efd