FDA 483 - QuVa Pharma, Inc. - December 08, 2016
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
On December 8, 2016, FDA investigators Stephen D. Brown and Lisa R. Jennings issued a Form 483 to QuVa Pharma, Inc., an outsourcing facility located at 5920 S General Bruce Dr Ste 100, Temple, TX 76502-5804. The inspection, conducted from November 28 to December 8, 2016, identified several deficiencies.
**Key Observations:**
1. **Aseptic Processing Area Cleaning and Disinfection:** The firm failed to perform effectiveness testing for disinfectants, such as [(b) (4)] and [(b) (4)], routinely used in ISO 5 [(b) (4)] areas. 2. **Environmental Monitoring:** The firm did not continuously record pressure differentials between ISO 5 [(b) (4)] and ISO 7 Clean Rooms during aseptic drug production. Additionally, pressure gauges measuring the differential from the [(b) (4)] to the [(b) (4)] lacked an alarm system, violating SOP COR-SOP-T0-0002, Version 01, Section 4.2. 3. **Air Supply Filtration:** The "Air Flow Smoke Pattern Test" in the [(b) (4)] Certification Report for each [(b) (4)] lacked documentation to show it was performed under dynamic conditions. 4. **Drug Product Label
- Company
- QuVa Pharma, Inc.
- Inspection Date
- December 8, 2016
ID · e7c58cc6-8304-425b-9536-979bf8fd50f5
Full citation text and observation details available on the Dashboard.